OB1203 Certification Preparation, Setup, and Test Procedure
X0119984 Rev.1.0
Mar 11, 2021
Page 9
c. Connect the Flash programmer to the adapter board and the adapter board to the four pin E1 connector
on the RL78 evaluation board. The pinout of the connector is listed on the evaluation board. The pins
are:
○
Vcc (3.3 VDC)
○
Tool0
○
Reset
○
Ground
d. Flash the *.mot file on the RL78
5. Test Setup Calibration using the Reference Board
Run the calibration software on the OB1203SD-RL-EVK Reference board. It scrolls through four screens that
illustrate calibration values necessary for this process.
SpO2 / HR
LED current
ADC counts for IR and Red
Software version
Figure 11. Calibration Software Screens