OB1203 Certification Preparation, Setup, and Test Procedure
X0119984 Rev.1.0
Mar 11, 2021
Page 13
6.3
Crosstalk Correction Validation
A Whaleteq AECG100 simulator was used to validate Sp02 measurement performance.
The following graph shows that the cover window induces significant inaccuracy while the cross-coupling
correction factor corrected the measurement.
Figure 13. Before and After Correction
The results in the graph show that while cross-coupling through a cover window can have a large effect on
accuracy, the OB1203 software can successfully correct for any particular cover window inducing cross-
coupling.