OB1203 Certification Preparation, Setup, and Test Procedure
X0119984 Rev.1.0
Mar 11, 2021
Page 2
1. Test Purpose
Customers who desire to obtain regulatory certification for their product equipped with Renesas’ OB1203 will
need to perform human subject testing. Since these tests are time consuming and complex, it is best to prepare
by performing pre-setup and pre-tests dry runs.
This document describes the following required steps for a successful completion:
1. Collecting the required equipment
2. Preparation of measurement
3. Test setup calibration
4. Test data collection
5. R-curve generation based on data collected
6. Validation test using the test setup
7. Creation of measurement threshold
2. Required Equipment
■
WhaleTeq AECG100 with 940nm test head
●
WhaleTeq AECG100 GUI installed
■
OB1203SD-RL-EVK Heart Rate, SpO2, and Respiration Rate Evaluation kit (reference board).
●
With installed calibration software
●
Renesas E2 or E2 Lite programmer including programming software
●
Renesas E2 or E2 Lite to OB1203SD-RL-EVK programming adapter board
■
Renesas opto-mechanical test adapter with inserted Aluminum reflector cone
■
The device under test – DUT
3. Preparation of Test Equipment
3.1
WhaleTeq AECG100 Hardware and Software Setup
Follow the
to complete the setup and proceed to the next step.
Figure 1. SP02 Test Setup with WhaleTeq AECG100