OB1203 Certification Preparation, Setup, and Test Procedure
X0119984 Rev.1.0
Mar 11, 2021
Page 6
4. Reference Board Setup and Preparation
To simplify this step, it is recommended to use an OB1203SD-RL-EVK board from Renesas. Please contact
Renesas support to request.
Each OB1203SD-RL-EVK is equipped with test mode software. In order to switch between “normal mode” and
“test mode”, push the wake button again until “Run mode change” appears on the display. The test software
supports printing test values on the display for easy post-processing.
This OB1203 test software is identical to the Renesas provided version running on the DUT except for the
following:
Parameter in Header
File OB1203.h
SpO2 / HR
Algorithm
Test Software
Comment
Light level ADC
THRESH_COUNT
196 k
32 k
Required threshold of light
coming from the WhaleTeq to
start algorithm
PS_THRESHOLD_HI to
start algorithm
15 k
200
Required threshold of light
coming from the WhaleTeq to
start algorithm
BIO_THRESHOLD
8 k
2 k
Display
SpO2 / HR
SpO2 / HR
ADC counts for IR and Red