Attaching the mirror housing:
Remove the back of the mirror housing, after loosening the four
corner fasteners by turning each fastener counterclockwise 1/4
turn.
Fasten the housing to the microscope stand with the captive
allen screw inside the housing (about a third of the way down in
the center) using a long allen wrench or other hex key tool. Do
this with care so no prism or filter is damaged.
Remove the protective covers and foam blocks for the sliding
mirror, rotating prism and oscillating mirror. Do not touch optical
parts with your fingers.
Insert the filter holder with IR-blocking filters in the filter holder
on the rear. The filter holder top is marked either X or H .
The filter holder X is used for working with the XBO 450W Xenon
lamp; for the HBO 200W mercury-vapor pressure lamp, use filter
holder H.
Replace the back panel of the mirror housing on the orientation
pin located above, and secure it with the four captive corner
fasteners by tightening them about 1/4-turn clockwise.
Be careful that the filters are not damaged.
Note! If the rear panel heat-blocking filters are not in place, do
not operate the xenon or mercury-vapor lamps, since the prisms
will be damaged by the heat from the lamp.
Mounting the lamp housings:
Remove the protective caps on the mirror housing and attach the
lamp housings with the screws at the back. See the description
of the lamp housing and the electrical connection on the lighting
equipment.
Zoom lighting:
The connection cable for the automatic zoom lighting is
connected between the microscope base and the power supply
for the 100W low-voltage halogen lamps.