22
Data Stream Output
Using Terminal application, a data stream can be viewed from the RS-232
output, connect RS-232/USB cable between the product and terminal/PC
and use the following setting to view the data stream.
Terminal Settings:
• Bits per second: 9600
• Data bits: 8
• Parity: none
• Stop bits: 1
47060100000209 = O
2
data (Probe 3) +20.9 % O
2
Reading
Decimal Place 0-3
Positive (+) = 0, Negative (-) = 1
°C= 01, °F= 02, PPM= 19,
%O
2
= 06, %RH= 04
%RH data (probe 1) = 1
Temp./%RH data (probe 1) = 2
Dew point data (probe 1) = 3
Wet bulb data (probe 1) = 4
CO
2
data (probe 2) = 5
Temp./CO
2
data (probe 2) = 6
O
2
data (probe 3) = 7
Temp./O
2
data (probe 3) = 8
CO data (probe 3) = 7
Temp./CO data (probe 3) = 8
4
47060100000209 = Données de O
2
(sonde 3) +20.9% O
2
Lecture
Position décimale 0-3
Positif (+) = 0, Negatif (-) = 1
°C = 01, °F = 02, PPM = 19,
%O
2
= 06, %RH = 04
Données d'humidité relative en % (sonde 1) = 1
Données de température/d’humidité
relative en % (sonde 1) = 2
Données du point de rosée (sonde 1) = 3
Données du thermomètre à
réservoir mouillé (sonde 1) = 4
Données de CO
2
(sonde 2) = 5
Données de température/CO
2
(sonde 2) = 6
Données de O
2
(sonde 3) = 7
Données de température/O
2
(sonde 3) = 8
Données de CO (sonde 3) = 7
Données de température/CO (sonde 3) = 8
4
3.5mm Terminal to RS-232 Serial Connection
RS-232
Mono Jack
Pin 2 (RX)
Pin 4 (DTX)
Sleeve
Tip