5. Select which Sensor Sensitivity you want to calibrate at:
Low Light Only: This option displays when the current Sensor Sensitivity is Low Light.
Standard Only: This option displays when the current Sensor Sensitivity is Standard.
Standard and Low Light: This option always displays.
Figure: Select Sensor Sensitivity for Auto Calibration
6. Select OK on the dialog box that opens. The system initializes and then makes multiple passes through the
calibration phases.
7. When the Calibration Successful dialog displays, select OK to complete the process.
NOTE: When you change the exposure setting, the E indicator turns yellow and it may take up to 8 seconds for the
calibration map to load.