5.2  Module 2 - S

econdary

 f

unction

 p

araMeterS

 (

2-SEC

)

SEL

Max Capture 

Delay Time

Max Display 

Enable

Min Display 

Enable

Access Code 

For Service 

Operations

Min Capture 

Delay TIme

Factory 

Service 

Operations

2-SEC

HI-En

HI-t

LO-En

LO-t

FCS

CodE

Pro

PARAMETER MENU

MIN DISPLAY ENABLE

YES

NO

2.0

HI-t

NO

LO-En

NO

FCS

MAX CAPTURE DELAY TIME

When the Input Display is above the present MAX value for the entered 

delay time, the meter will capture that display value as the new MAX reading. 

A delay time helps to avoid false captures of sudden short spikes.

2.0

LO-t

MIN CAPTURE DELAY TIME

When the Input Display is below the present MIN value for the entered delay  

time, the meter will capture that display value as the new MIN reading. A delay 

time helps to avoid false captures of sudden short spikes.

0.0

 

to

 

999.9

 

seconds

NO

HI-En

MAX DISPLAY ENABLE

0.0

 

to

 

999.9

 

seconds

Select 

YES

 to perform either of the Factory Service Operations shown below.

FACTORY SERVICE OPERATIONS

yES

NO

YES

NO

Enables the Maximum Display Capture capability.

Enables the Minimum Display Capture capability.

The CUB5TC uses stored voltage calibration and cold 

junction temperature values to provide accurate temperature 

and voltage measurements. Over time, the electrical 

characteristics of the components inside the meter could 

slowly change. The result is that the stored calibration values may no longer 

accurately define the input circuit. For most applications, recalibration every 1 

to 2 years should be sufficient.

Calibration of the CUB5TC involves a voltage calibration and a cold 

junction calibration. It is recommended that both calibrations be performed. 

The voltage calibration MUST precede the cold junction calibration. Allow 30 

minute warm up before performing any calibration related procedure. The 

following procedures should be performed at an ambient temperature of 15 to 

35 °C (59 to 95 °F).

Calibration should only be performed by individuals experienced in 

calibrating electronic equipment.

CAUTION

: The accuracy of the calibration equipment will directly affect the 

accuracy of the CUB5TC.

Input Voltage Calibration

1. Connect a precision DC voltage source with an accuracy of 0.01% or better 

to the TC+ (positive) and the TC- (negative) terminals of the CUB5TC. Set 

the output of the voltage source to zero.

2. With the display at 

CodE 48

, press and hold the 

SEL

 button for 2 seconds. Unit 

will display 

CAL NO

.

3. Press the 

RST

 button to select 

INP

.

4. Press the 

SEL

 button. Display reads 

0.0u

.

5. With the voltage source set to zero, press 

SEL

. Display reads 

CALC

 for about 

eight seconds.

6. When display reads 

60.0u

, apply 60.000 mV input signal. Press 

SEL

. Display 

reads 

CALC

 for about eight seconds.

7. When display reads 

CAL NO

, press 

SEL

 twice to exit Module 2 and return to 

the normal display mode.

8. Proceed to Cold Junction Calibration.

Cold Junction Calibration

1. 

Install all option cards needed for your application and the rear cover, 

or invalid results will occur.

2. The ambient temperature must be within 20°C to 30°C.

3. Connect a thermocouple (types T, E, J, K, or N only) with an accuracy of 1°C 

or better to the meter.

4. Enter programming mode and verify the following settings in Module 1:

 

tYPE

 = thermocouple type connected to the unit

 

CJC

 = 

YES

;     

SCALE

 = 

°C

;     

dECPt

 = 

0.0

;     

OFSEt

 = 

0

5. Place the thermocouple in close thermal contact to a reference thermometer 

probe. (Use a reference thermometer with an accuracy of 0.25°C or better.) 

The two probes should be shielded from air movement and allowed sufficient 

time to equalize in temperature. (A calibration bath of known temperature 

could be used in place of the thermometer.)

6. Compare the unit display with the reference temperature indicator (or 

calibration bath). If a difference of more than ±1.0 °C exists, note the 

difference (CJ error) and continue with cold junction calibration.

 

CJ Error = Reference Temperature - Unit Display.

7. Enter programming mode. Step through Module 2 to the Service Access 

Code parameter and select 

CodE 48

. Press and hold the 

SEL

 button until the 

unit displays 

CAL NO

. Press the 

RST

 button to select 

CJC

.

8. Press 

SEL

. Display reads 

CJC

 followed by the current cold junction value. 

Calculate a new cold junction value as follows:

 

New cold junction = Current cold ju CJ Error (noted above)

9. Press 

RST

 and set the display to the new cold junction value. Press and hold 

SEL

. Display reads 

CALC

 for about four seconds and then returns to 

CAL NO

.

10. Press 

SEL

 twice to exit calibration and return to the normal display mode. 

Verify the input reading is correct. If not, repeat steps 6 through 10.

CALIBRATION

48

CodE

RESTORE FACTORY DEFAULT SETTINGS

Entering Code 66 will overwrite all user settings with 

the factory settings. The meter will display 

rESEt

 and then 

return to 

CodE 00

. Press 

SEL

 button to exit the module. 

Pressing both the 

SEL 

and the 

RST 

button on power-up 

will also load the factory settings and display 

rESEt

. This allows operation in 

the event of a memory failure or corruted data.

66

CodE

Содержание CUB 5TC-K

Страница 1: ...ure personal safety and to prevent damage to either the device or equipment connected to it Do not use these products to replace proper safety interlocking No software based device or any other solid...

Страница 2: ...nstallation Guidelines for additional information 8 MEMORY Nonvolatile E2PROM memory retains all programming parameters and max min values when power is removed 9 CONNECTIONS Wire clamping screw termi...

Страница 3: ...nstalling the Meter INSTALLATION The meter meets NEMA 4X IP65 requirements when properly installed The unit is intended to be mounted into an enclosed panel Prepare the panel cutout to the dimensions...

Страница 4: ...tive The following EMI suppression devices or equivalent are recommended Fair Rite part number 0443167251 RLC part number FCOR0000 Line Filters for input power cables Schaffner FN2010 1 07 Red Lion Co...

Страница 5: ...order to preserve the safety of the meter application the power input common and the sensor input common must be suitably isolated from hazardous live earth referenced voltages or input common must b...

Страница 6: ...ges to memory and return the meter to the Display Mode If power loss occurs before returning to the Display Mode verify recent parameter changes PROGRAMMING TIPS It is recommended to start with Module...

Страница 7: ...SELECTION VOLt tc n N K tc b tc S tc K B S J tc j E tc E T tc t 0 OFSEt DISPLAY OFFSET VALUE The temperature display can be corrected with an offset value This can be used to compensate for probe erro...

Страница 8: ...it will display CAL NO 3 Press the RST button to select INP 4 Press the SEL button Display reads 0 0u 5 With the voltage source set to zero press SEL Display reads CALC for about eight seconds 6 When...

Страница 9: ...rogramming Immediate Access Entering Code 222 allows access regardless of security code 5 3 MODULE 3 Display and Front Panel Button Parameters 3 dSP SEL Pro Front Panel Display Select Enable Display U...

Страница 10: ...anced Hys LO bL OUTPUT STATE OFF ON Hys SP SP Hys OFF TRIGGER POINTS High Acting Unbalanced Hys HI Ub 2 HYS n HYSTERESIS VALUE 1 to 59999 Enter desired hysteresis value See Setpoint Output Figures for...

Страница 11: ...the output operates normally per the Setpoint Action and OutputReset Action NO Stb n STANDBY OPERATION YES NO When the user input or RST button is activated momentary action the corresponding on outpu...

Страница 12: ...gramming Security Code CodE Display Decimal Point Thermocouple Type tYPE Cold Junction Compensation CJC 1 INP Temperature Scale SCALE dECPt USrIN User Input Function Filter Setting Display Offset Valu...

Отзывы: