Cold Junction

Compensation

Temperature 

Scale

Display 

Decimal Point

Filter Setting

User Input 

Assignment

Thermocouple 

Type

SEL

tYPE

CJC

SCALE

dECPt

OFSEt

FILtr

bANd

U-ASN

1-INP

Pro

User Input 

Function

Display Offset 

Value

USrIN

Filter Band

7

5.1  Module 1 - S

iGnal

 i

nput

 p

araMeterS

 (

1-INP

)

PARAMETER MENU

tc-J

tYPE

THERMOCOUPLE TYPE

0

dECPt

0.0

0

DISPLAY DECIMAL POINT

Select the decimal point location for the desired display resolution. This 

selection applies for the Input, MAX and MIN displays. This parameter does 

not appear if 

tYPE

 = 

VOLt

 or for types R, S or B thermocouples which have a 

fixed 1 degree resolution.

1

FILtr

FILTER SETTING

If the displayed temperature is difficult to read due to small process 

variations or noise, increased levels of filtering will help to stabilize the display. 

Software filtering effectively combines a fraction of the current input reading 

with a fraction of the previous displayed reading to generate the new display.

0,1 2 3

10

bANd

FILTER BAND

The filter will adapt to variations in the input signal. When the variation 

exceeds the input filter band value, the filter disengages. When the variation 

becomes less than the band value, the filter engages again. This allows for a 

stable readout, but permits the display to settle rapidly after a large process 

change. The value of the band is in display units, independent of the Display 

Decimal Point position. A band setting of ‘0’ keeps the filter permanently 

engaged at the filter level selected above.

00

 

to

 

199

 

display units

Select the thermocouple type used for the application. The appropriate curve 

will be automatically loaded for the selected type.

Selecting 

VOLt

 displays the millivolt input signal with 10 µV resolution.

°F

SCALE

TEMPERATURE SCALE

Select the temperature scale. This selection applies for the Input, MAX and 

MIN displays. This parameter does not appear if 

tYPE

 = 

VOLt

.

°C

°F

R

tc-r

SELECTION

TC TYPE

TC TYPE

SELECTION

VOLt

tc-n

N

K

tc-b

tc-S

tc-K

B

S

J

tc-j

E

tc-E

T

tc-t

0

OFSEt

DISPLAY OFFSET VALUE

The temperature display can be corrected with an offset value. This can be 

used to compensate for probe errors, errors due to variances in probe placement 

or adjusting the readout to a reference thermometer.

-999

 

to

 

9999

This parameter enables or disables internal cold junction compensation. For 

most applications, cold junction compensation should be enabled (

yES

). This 

parameter does not appear if 

tYPE

 = 

VOLt

.

COLD JUNCTION COMPENSATION

NO

yES

yES

CJC

USER INPUT FUNCTION

NO

USrIN

Setpoint 1 and 2 Reset

Setpoint 1 Reset
Setpoint 2 Reset

Print and Reset

Reset both setpoint 1 and 2 outputs.

Resets setpoint 1 output.
Resets setpoint 2 output.

Same as Print Request followed by a 

momentary reset of the assigned value(s).

rSt12

rSt-1
rSt-2

P-r5t

Print Request

Serial transmit of the active parameters 

selected in the Print Options menu 

(Module 5).

Print

Select the value(s) to which the User Input Function is assigned. The User 

Input Assignment only applies if a selection of reset, display hold, or print and 

reset is selected in the User Input Function menu.

USER INPUT ASSIGNMENT

dSP

U-ASN

Filter values represent no filtering (0), up to heavy filtering (3). A value of 1 

for the filter uses 1/4 of the new input and 3/4 of the previous display to 

generate the new display. A filter value of 2 uses 1/8 new and 7/8 previous. A 

filter value of 3 uses 1/16 new and 15/16 previous.

dSP

HI-LO 

LO

HI

 MODE

Reset (Edge triggered)

Program Mode Lock-out

No Function

DESCRIPTION

DISPLAY

Display Select  

(Edge Triggered)

Advance once for each activation.

d-SEL

Backlight Color  

(Edge Triggered)

Change backlight color with each 

activation (backlight version only).

COLOr

Display Intensity Level  

(Edge Triggered)

Increase intensity one level for each 

activation (backlight version only).

d-LEV

Display Hold

Holds the assigned display, but all other 

meter functions continue as long as 

activated (maintained action).

Resets the assigned value(s) to the 

current input value.

rESEt

d-HLd

See Programming Mode Access chart 

(Module 3).

P-Loc

User Input disabled.

NO

Содержание CUB 5TC-K

Страница 1: ...ure personal safety and to prevent damage to either the device or equipment connected to it Do not use these products to replace proper safety interlocking No software based device or any other solid...

Страница 2: ...nstallation Guidelines for additional information 8 MEMORY Nonvolatile E2PROM memory retains all programming parameters and max min values when power is removed 9 CONNECTIONS Wire clamping screw termi...

Страница 3: ...nstalling the Meter INSTALLATION The meter meets NEMA 4X IP65 requirements when properly installed The unit is intended to be mounted into an enclosed panel Prepare the panel cutout to the dimensions...

Страница 4: ...tive The following EMI suppression devices or equivalent are recommended Fair Rite part number 0443167251 RLC part number FCOR0000 Line Filters for input power cables Schaffner FN2010 1 07 Red Lion Co...

Страница 5: ...order to preserve the safety of the meter application the power input common and the sensor input common must be suitably isolated from hazardous live earth referenced voltages or input common must b...

Страница 6: ...ges to memory and return the meter to the Display Mode If power loss occurs before returning to the Display Mode verify recent parameter changes PROGRAMMING TIPS It is recommended to start with Module...

Страница 7: ...SELECTION VOLt tc n N K tc b tc S tc K B S J tc j E tc E T tc t 0 OFSEt DISPLAY OFFSET VALUE The temperature display can be corrected with an offset value This can be used to compensate for probe erro...

Страница 8: ...it will display CAL NO 3 Press the RST button to select INP 4 Press the SEL button Display reads 0 0u 5 With the voltage source set to zero press SEL Display reads CALC for about eight seconds 6 When...

Страница 9: ...rogramming Immediate Access Entering Code 222 allows access regardless of security code 5 3 MODULE 3 Display and Front Panel Button Parameters 3 dSP SEL Pro Front Panel Display Select Enable Display U...

Страница 10: ...anced Hys LO bL OUTPUT STATE OFF ON Hys SP SP Hys OFF TRIGGER POINTS High Acting Unbalanced Hys HI Ub 2 HYS n HYSTERESIS VALUE 1 to 59999 Enter desired hysteresis value See Setpoint Output Figures for...

Страница 11: ...the output operates normally per the Setpoint Action and OutputReset Action NO Stb n STANDBY OPERATION YES NO When the user input or RST button is activated momentary action the corresponding on outpu...

Страница 12: ...gramming Security Code CodE Display Decimal Point Thermocouple Type tYPE Cold Junction Compensation CJC 1 INP Temperature Scale SCALE dECPt USrIN User Input Function Filter Setting Display Offset Valu...

Отзывы: