CMM 100 :
INSTALLATION, CALIBRATION AND OPERATION
15
5. Data Logging
The instrument continuously outputs a line of
data every second. A 'Wizard Upload Lead'
(Pt No. S/CB/327-1-030) is required, and
suitable data capture software must be
installed on your PC. This software is not
provided by RDS.
Each time you press the
CAL
button, the
instrument transmits a header line and the
'Sample ID' (line count) is reset to zero.
5.1
Setup
Connect the 'Wizard Upload Lead' to the 4-
way header 'PL1' in the 'Terminator' junction
box. This lead is then connected to an
RS232 serial port on a PC or laptop
computer.
If a serial port is not available, a Serial-to-USB
converter can be used (available from any
computer store).
Port settings required are,
Baud Rate:
4800
Data Bits:
8
Parity:
None
Handshake:
OFF
Mode:
Free
5.2
Data Dictionary
Typical data line:
1, A, 90.00, 2.35, 3.80, +38.4, 20.7, 19.6
Sample ID
Channel
Average Filter Ratio
Moisture Sensor Input (V)
Temp. Sensor Input (V)
Calculated Temp. (deg C)
Calculated Moisture Content (%)
Calculated Temp. compensated Moisture Content ( % )
Connect 'Wizard
Upload Lead'
here