CMM 100 :
INSTALLATION, CALIBRATION AND OPERATION
10
3.
Calibration Mode
3.1
Navigating the Calibration Mode
1.
Press and hold the button as you switch the instrument on, and
continue to hold for 5 seconds until the instrument enters the calibration
mode. The instrument displays "
CAL
" then "
ALAr
" (for the channel last
selected in Normal Operating Mode), then the current moisture alarm
threshold setting (
xx.xx
).
2
.
Press the button to select the required channel. The indicator
shows which channel is selected.
3.
Press the button to cycle to the required setting for the selected
channel.
Factory Default
settings
Channel (Grain Type)
A
B
C
D
E
F
M.C. Alarm ("
LO
")
Threshold ("
HIGH
")
("
ALAr"
)
5.00
25.00
5.00
25.00
5.00
25.00
5.00
25.00
5.00
25.00
5.00
25.00
High/Low Level Alarm
option
("HIGH"/"LO")
Factor A
("
A.FAc
")
00.00 00.00
00.00
00.00
00.00
00.00
Factor B
("
B.FAc
")
7.04
3.75
5.52
2.12
4.01
7.91
Factor C
("
C.FAc
")
6.60
12.20
10.40
12.00
7.10
2.00
Low Grain Voltage
Threshold
("
UOLt
")
0.00
0.00
0.00
0.00
0.00
0.00
Smoothing Factor
("
FILt
")
99.5
99.5
99.5
99.5
99.5
99.5
The default settings are for grain types:
A = Wheat, B = Barley, C= Oats, D = Oilseed Rape, E = Beans
F = Maize