CMM 100 :
INSTALLATION, CALIBRATION AND OPERATION
12
3.3
Factor A ("A.FAc") / Factor B ("B.FAc")
May require changing if harvesting / drying crops other than the default grain
types.
To edit - Ref. section 3.2
.
3.4
Factor C ("C.FAc")
This setting is the Moisture Correction Factor, required to match the moisture
content reading of the instrument to that of a reference moisture meter. It can
be changed within the calibration mode, or in the Normal Operating Mode as
described (ref. section 2.3.1).
To edit - Ref. section 3.2
.
3.5
Moisture Content Alarm ("ALAr")
This setting is either the minimum or maximum moisture content (%) below
or above which, an alarm warns the operator.
The instrument is designed either for combine installation where the high
alarm is enabled, or for dryer installation where the low alarm is enabled. The
default setting is for a high alarm.
The factory-default setting for all channels is 5% moisture content (low alarm)
and 25% (high alarm). It can only be changed within the calibration mode.
To edit alarm settings - Ref. section 2.5 - 2.6
.
3.6
Low Grain Voltage Threshold ("UOLt")
This setting (Volts) determines when the instrument displays "
Ch.A-F
" to
indicate insufficient grain flow over the sensor. It is possible that this value
may need to be adjusted for each grain type.
To edit - Ref. section 3.2
.