Cobra-Tac User's Guide
page 46
RD Instruments
f. Connect the waterproof power cable connector after charging. The
charger sockets are not waterproof and must be capped when connected
to the Cobra-Tac.
6.3
Charging Batteries Safely
Follow all Charging Safety warnings while charging batteries.
CAUTION.
When charging batteries, ventilate the charging area. A
buildup of hydrogen and oxygen in the battery or in the charging area can
create an explosion hazard.
CAUTION.
If the battery feels hot to the touch during charging, STOP.
Allow the battery to cool before charging again. A battery that is too hot
can explode.
CAUTION.
Use the provided charger cord. Unplug the charger or turn it
off before you disconnect the cable, this will cut down on the chance of
sparks.
CAUTION.
ABSOLUTELY NO SMOKING, SPARKS, OR FLAMES
AROUND CHARGING BATTERIES. Charging gives off hydrogen and
oxygen, which can explode if ignited.
6.4 Battery
Tips
•
Recharge the battery after each use.
•
Do not over discharge the battery; you will shorten the battery
life. Turn the Cobra-Tac off when not in use.
•
Store the battery fully charged, and recharge the pack for a few
hours each month.
•
Do not disassemble the battery.
NOTE.
Replace the battery if the measured voltage of the battery is less
than 11 volts after charging the battery for 5 hours or if the battery has
been used for 500 charge/discharge cycles.
6.5
I/O and Power Cable
The I/O connector (on the Cobra-Tac
) and the I/O cable and power cable
use molded connectors.
The cable should be disconnected with a straight outward motion only. Use
the following procedure to disconnect the cable.
a. Release the retaining strap by pulling it over the connector.
Содержание Cobra-Tac