Command is designed for rough calibration («zero installation») in the process of FLS
installation. If the length of probe is within (150…710) mm, «zero installation» is not required.
Command format:
Drift, byte
Size of field, byte
Value
Description
0
1
31h
Prefix
+1
1
00h…FFh
Network address of
acceptor
+2
1
45h
Command code
+3
1
00h…3Fh
Value of parameter of
rough calibration
+4
1
00h…FFh
CRC8
Answer format:
Drift, byte
Size of field, byte
Value
Description
0
1
3Eh
Prefix
+1
1
00h…FFh
Network address of
sender
+2
1
45h
Command code
+3
1
00h or 01h
Return code: 00h -
command is complied
successfully, 01h -
Error
+4
1
00h…FFh
CRC8
A2.3.2.2.2
Exact calibration (command 49h).
Command is designed for exact calibration («zero installation») in the process of FLS
installation. If the length of probe is within (150…710) mm, «zero installation» is not required
Command format:
Drift, byte
Size of field, byte
Value
Description
0
1
31h
Prefix
+1
1
00h…FFh
Network address of
acceptor
+2
1
49h
Command code
+3
2
0000h…FFFFh
Value of parameter of
exact calibration
+5
1
00h…FFh
CRC8
Answer format:
Drift, byte
Size of field, byte
Value
Description
0
1
3Eh
Prefix
+1
1
00h…FFh
Network address of
sender
+2
1
49h
Command code
+3
1
00h or 01h
Return code: 00h -
command is complied
successfully, 01h -
Error