RBR#0008819revA

 –  

14

1.
2.
3.
4.
5.

1.
2.

Most experienced users of oceanographic equipment replace all O-rings before every deployment as a matter of 
routine. The cost of an O-ring is negligible compared with the cost of the instrument and its deployment. Therefore, 
routine replacement of O-rings is cheap insurance.

O-rings: #2-225 N70 O-ring (RBR#0002408), #8-225 N90 backup ring (RBR#0009115)

6.4.2  Servicing the RBRquartz BPR|zero waste oil reservoir

Overview

The RBR

quartz

3

 BPR|zero utilizes a

 

Paroscientific Digiquartz® sensor which is filled with a buffer oil and is connected to 

the environment via a tube.

 

The difference between an RBR

quartz

3

 BPR and BPR|zero is that the |zero makes use of a 

valve to change the 

Paroscientific Digiquartz® sensor or sensors to measure the pressure inside the housing of the 

instrument, which is monitored by a barometer in order to correct for the drift of the sensors over time. This actuation 
of this valve results in a small volume of buffer oil transferred into the pressure housing of the instrument. The 
transferred buffer oil is stored in a reservoir that will need to be cleaned periodically. The instructions in this section will 
describe how to open this reservoir and replace the absorbent towel.

Materials required:

Protective gloves
Eye protection
Replacement absorbent towels*
O-ring*
Silicone grease*

*Provided in the maintenance kit

Opening the reservoir

Open the battery end cap as described in 

Opening and closing the RBRquartz BPR|zero

Remove the umbilical from the battery end cap by pushing down on the latch (located on the left-
hand side of the connector) and pulling the connector down. Do not pull from the wires attached to 
the connector, as this could damage them.

                        

Содержание quartz3 BPR zero

Страница 1: ...rbr global com INSTRUMENT GUIDE BPR zero...

Страница 2: ...9 6 2 Changing the batteries RBRquartz BPR zero 11 6 2 1 Steps 11 6 3 Installing desiccant RBRquartz BPR zero 12 6 4 Maintenance and repairs 12 6 4 1 O rings RBRquartz BPR zero 12 Inspecting O rings 1...

Страница 3: ...0008819revA 3 2 Introduction You can access the RBR instrument guide on the USB data stick provided when you purchase a logger from the Help menu in Ruskin and on the RBR web site at www rbr global c...

Страница 4: ...RBR 0008819revA 4 3 Revision history Revision No Release Date Notes A 04 May 2020 Original...

Страница 5: ...by us The liability of RBR Ltd extends only to the replacement cost of the instrument The customer will bear all costs of shipment to us for repair all other costs including return shipment will be b...

Страница 6: ...rements are critical The RBRquartz BPR zero may be supplied with external power from RBRfermata power or RBRcervata power and memory canisters or cabled to an observatory for external power and realti...

Страница 7: ...35 C The instrument has 8 X AA batteries installed however this is intended only for purposes of sustaining the instrument during external power interruptions of a few minutes While the Paros will con...

Страница 8: ...2000 4000 7000 dbar Initial accuracy 0 01 FS full scale Resolution 10ppb at 16Hz sampling rate Accelerometer optional Range 3g Resolution 100ng The full mechanical drawing is available in Appendix A...

Страница 9: ...shown below Only the battery end cap needs to be opened to access the data connector batteries desiccant and waste reservoir To open the logger battery end cap the two M6 screws will need to be remov...

Страница 10: ...il the battery end cap is mated fully to the housing Remember to keep the O rings clean and avoid scratching the O ring mating surfaces Carefully inspect the O rings as described in the Inspecting the...

Страница 11: ...ent during external power interruptions of a few minutes While the Paros will continue to run and be measured the valve will not operate 6 2 1 Steps Remove the battery end cap from the logger see Open...

Страница 12: ...desiccant capsules can be opened the desiccant removed and replaced or dried 6 4 Maintenance and repairs 6 4 1 O rings RBRquartz BPR zero The single most important item of maintenance on any RBR subme...

Страница 13: ...bishment please contact us for advice Replacing an O ring Lever the O ring from its groove Use a soft plastic or wood tool do not use a metal screwdriver or any other tool which may scratch the surfac...

Страница 14: ...easure the pressure inside the housing of the instrument which is monitored by a barometer in order to correct for the drift of the sensors over time This actuation of this valve results in a small vo...

Страница 15: ...towel around the reservoir in a clockwise manner when viewed as below The towel should only wrap around the reservoir once any additional layers will make it difficult to insert 8 While holding the to...

Страница 16: ...closely as possible to the current market value based on the original purchase price less depreciation Please contact RBR if you have any questions about the instrument value Send the unit using your...

Страница 17: ...RBR 0008819revA 17 7 Appendix 7 1 Appendix A RBRquartzBPR zero Product Drawing...

Страница 18: ...RBR 0008819revA 18...

Страница 19: ...RBR 0008819revA 19 7 2 Appendix B RBRquartzBPR zero Service Drawing...

Страница 20: ...RBR 0008819revA 20 7 3 Appendix C RBRquartzBPR zero Connector Details...

Отзывы: