RBR#0008819revA
–
7
Power:
Requires external power**
External power:
9.5-30VDC
Communication:
USB-C, RS-232/485, Ethernet***
Other
Storage:
240M readings
Clock drift:
±60 seconds/year
Integration time
760ms (Single Paros)
720ms (Dual Paros)
*Calibrated temperature range is -5°C to 35°C
**The instrument has 8 X AA batteries installed, h
owever, this is intended
only
for purposes of sustaining the instrument
during external power interruptions of a few minutes. While the Paros will continue to run and be measured, the valve
will not operate.
***Connector diagram and pinout available in
Appendix C - RBRquartzBPR|zero Connector Details
Marine temperature (standard)
Range:
-5 to 35°C
Accuracy:
±0.002°C
Time constant:
~3 minutes
Typical stability:
±0.002°C/year