background image

10

MagCore® HF48

Press Start key in the main functional window 

Select 24 or 48 test channels 

Select Cartridge number 

Cartridge number is marked on the Cartridge you use.

Please reconfirm the number you selected and select all 
parameters 

Confirm that the icon              in the status bar is in normal status  If 
the icon shows  , the Start key will be at “inactive” status and the 
operation is disabled  Please follow the display portion to remove 
error to solve  (Refer to page 8 for warning marks)

When you finish input and press Start key, the system will show 
a confirmation window, on which there are the parameters you 
input  Please press Yes to operate the system if the parameters are 
not incorrect  Please press No to modify incorrect parameters if you 
want to modify 

You may watch the procedure status and remained time of the 
current operation on the control panel  The system will display 
Complete page after completion 

━ Start

Start

Maintenance

Load program

Setup

Status

2012.06.07

12:12:12

Start

Stop

Save

Back

Load

Program

Start

24 Test

48 Test

Cartridge Code

Dnase Treatmen

Elution Buffer

Sample Volume

Elution Volume

A

B

D

A

B

E

Start

Cartridge Code

DNase I Treatment

Sample Volume

Elution Volume

Please select Optical Detrction

Optical Detection

Save

Back

Load

Program

101 MagCore Genomic DNA Whole Blood Kit (Speedy installation)

Start

Yes

No

48

Test

Test : 

Cartridge Code : 

Remove DNA

Sample Volumn : 

Elution Volumn :

Elution Buffer : 

48
105 Plasma DNA Extraction Kit (1.2 ml)
---
200 µl
150 µl
---

Start

Stop

Save

Back

Load

Program

Start

24 Test

48 Test

Cartridge Code

Dnase Treatmen

Elution Buffer

Sample Volume

Elution Volume

A

B

D

A

B

E

Please select Channel

Press START to Proceed

105 Plasma DNA Extraction Kit (1.2 ml)

200 µl

150 µl

Min

90

Processing

Stop

Save

Back

Load

Program

Heater B 68°C /  154.4 °F   |   Heater A 68°C /  154.4 °F

Heating

Test : 

Cartridge Code : 

Remove DNA

Sample Volumn : 

Elution Volumn :

Elution Buffer : 

48
105 Plasma DNA Extraction Kit (1.2 ml)
---
200 µl
150 µl
---

A

B

D

A

B

E

Min

90

Processing

Stop

Save

Back

Load

Program

Heater B 68°C /  154.4 °F   |   Heater A 68°C /  154.4 °F

Heating

Test : 

Cartridge Code : 

Remove DNA

Sample Volumn : 

Elution Volumn :

Elution Buffer : 

48
105 Plasma DNA Extraction Kit (1.2 ml)
---
200 µl
150 µl
---

A

B

D

A

B

E

Remained time for system operation

Heater status 

Display current operation status 

Display selected parameter items 

 Emergency Stop and Cancel keys  Please 
refer to the content on page 9 

Next Step

Next Step

Next Step

Next Step

 It is recommended not to open the 

viewing window or feeding door to 

prevent from poor extraction result in 

system operation if there is no emergency.

Complete

Back

Start

Содержание MagCore HF48

Страница 1: ...llCartridge 7 TouchPanelandUserInterface 8 MainMenu 8 PanelFunctionalArea 8 EmergencyStopandWarning 9 Start 10 LoadProgram 11 SavePrograms 12 DeletePrograms 12 UpdateProgramFiles 13 EasyMaintenance 13...

Страница 2: ...11 GroundingorPolarization Thisproductmaybeequippedwithapolarized alternating current line plug a plug having one blade wider than the other This plug will fit into the power outlet only one way This...

Страница 3: ...99 inches NetWeight 250kg 551 25Lb Specification OperationParameters OperationEnvironment Applications Whole Blood 200 400ul Viral Nucleic Acid DNA RNA Tissue Genomic DNA Plant Genomic DNA Bacteria Ki...

Страница 4: ...50 Tip TipHolderSetX 50 Elution Carrier X 1 CartridgeCarrierB X 1 CartridgeCarrierA X 1 TipCarrierB X1 TipCarrierA X1 PowerCord X1 Lubricant X1 O RingX 24 Systemtesttoolsshouldbekeptincleanplacesandth...

Страница 5: ...n Area Shutter Fixed Stands four legs Fixed Bolt on the Top Cover of Part Chamber SystemOverview Cooling Fans Power Switch and Plug Power Switch red Light Switch green Part Chamber Maintenance Cover S...

Страница 6: ...e for system installationandventilationbeforeinstallation When installation the system please make sure the elevation difference by using level rail for measurement and fixed stands forfine tuning Sys...

Страница 7: ...egroundlevel andtightenfixedscrews Shearthetransportationbeltandreturnthetopcoverofpart chamber ShearthebeltsonRacks Plugthepowercordandturnonthemainpower Turnonthepower red andthesystemwillbeinitiali...

Страница 8: ...ect usemay causethesystem inoperablebecauseCarrier isnot detected InstallCartridge Install Tip and SampleTube B A InsertTipandSampleTubeintotheslotsofTipCarrier Afterthat placetheentiresetofTipCarrier...

Страница 9: ...rial number error Setup Systemtimesetup Start Stop Save Back Load Program Start 24 Test 48 Test DNase I Treatment Sample Volume Elution Volume Please select Cartridge Code Elution Buffer Optical Detec...

Страница 10: ...e RemoveDNA SampleVolumn ElutionVolumn ElutionBuffer 48 105PlasmaDNAExtractionKit 1 2ml 200 l 150 l A A B B E D Warning Abort Program Stopped Press ABORT to terminate the process or CONTIUNE to run Co...

Страница 11: ...Sample Volume Elution Volume Please select Optical Detrction Optical Detection Save Back Load Program 101 MagCore Genomic DNA Whole Blood Kit Speedy installation Start Yes No 48 Test Test CartridgeCo...

Страница 12: ...red Program 3 Stored Program 4 Stored Program 5 Stored Program 6 Stored Program 7 Stored Program 8 Load Program Yes No 48 Test CartridgeCode RemoveDNA SampleVolumn ElutionVolumn ElutionBuffer 105Plasm...

Страница 13: ...1 2ml 200 l 150 l Start Stop Save Back Load Program Start 24 Test 48 Test Cartridge Code Dnase Treatmen Elution Buffer Sample Volume Elution Volume A B D A B E Please select Channel Youmaydeleteyourco...

Страница 14: ...ack UV Sterilization Moving Test Heater Test Rejection Tip Piston Test O Ring Piercing Needles DC Motor Operate Update Program Files EasyMaintenanceisforuser smaintenanceitems includingsystemssuchas s...

Страница 15: ...ings Step 2 Applywithnewone Replace O Rings Clean Piercing Needles Start Stop O Ring Piercing Needles Select item and Press Start to run Back Finish Replace O Rings Step 3 Usesuppliedlubricatingoil to...

Страница 16: ...horizontalorthereisaloweringsituation pleasepress Finishtotestagain Iftheproblemhappenscontinuously please refertopage14toreplacenewO RingforproblematicPiston andtestthisitemagainafterreplacement Ifth...

Страница 17: ...Program Files HeaterTest Start Stop Heater Test Press START to run Back HeaterA 65 C 148 F Heater B 65 C 148 F Thisfunctionismainlyforthesystemtoperformrejectionafter emergencystoporoperationinterrup...

Страница 18: ...Update Program Files Start Stop 30min 60min UV Sterilization select Item and press START to run Back 30 00 Remaining Time Complete Back UVSterilization Openthetopcoverofthepartchamber ApplynewUVlamp S...

Страница 19: ...jection Tip Piston Test O Ring Piercing Needles DC Motor Operate Update Program Files Maintenance Please select Item Back UV Sterilization Moving Test Heater Test Rejection Tip Piston Test O Ring Pier...

Страница 20: ...0 Q W E R T Y U I O P Backspace Enter A S D F G H J K L Z X C V B N M System Timing Setting Start Maintenance Loadprogram Setup Status 2012 06 07 12 12 12 Time Setting Back Year 2012 24 h Hour 12 Minu...

Страница 21: ...p Operating Time Hr 00 Min 00 Sec 00 UI Software Ver Ver SUPER2012 0 1 PLC Software Ver Ver SUPER2012 0 1 Machine serial number Error history list UV Time Reset Status Back System Date 2012 12 12 Syst...

Страница 22: ...Choosethecorrespondingslotofyoursampleandscanyour sample scode afterthecompletioncodeisdisplayedonthe touchpanel placethesampleintotheslot ifyouhavemorethan onesample pleaserepeattheabovebarcodescanni...

Страница 23: ...ystem test and check liquid levels of each columnareatthesamehorizontal Liquidleveloversuctionfilter 1 Wrongsamplematerials 2 Wrongsampletubes 1 PleaserefertoMagCore kitmanual 2 Pleaseusetheprovidedsa...

Страница 24: ...TEL 886 2 8912 1200 FAX 886 2 8912 1300 3F No 132 Ln 235 BaoqiaoRd XindianDist NewTaipeiCity23145 TAIWAN http www rbcbioscience com...

Отзывы: