90
RBCBioscience
10
Y_Home Sensor Fail
[Causes]
When performing the motor initial, the Home sensor has no
response and the motor collides into a hard stop.
[Solutions]
(1) Check whether the cable has dropped off.
(2) Replace the sensor.
11
V_S
[Causes]
Manual V JOG+ operation exceeds the distance programmed
in the software.
[Solutions]
(1) Alarm Reset
(2) Manual V JOG- operation
12
V_SoftLimt-
[Causes]
Manual V JOG- operation exceeds the distance programmed
in the software.
[Solutions]
(1) Alarm Reset
(2) Manual V JOG+ operation
13
V_Encode no change
[Causes]
No signal response from the Encoder.
[Solutions]
(1) Replace the cable connected to the Encoder.
(2) Replace the motor module.
14
V_Position Over 60 Pulse
[Causes]
The position command coordinate is vastly different to the
Encoder coordinate.
[Solutions]
(1) Check whether the terminal of the PLC Y00 Cable has
dropped off.
(2) Replace the cable connected to the power of the motor.
(3) Replace the main board.
(4) Replace the motor module.
15
V_Home Sensor Fail
[Causes]
When performing the motor initial, the Home sensor has no
response and the motor collides with hard stop.
[Solutions]
(1) Check whether the cable has dropped off.
(2) Replace the sensor.
16
X_S
[Causes]
Manual X JOG+ operation exceeds the distance of the
software.
[Solutions]
(1) Alarm Reset
(2) Manual X JOG- operation
Code No. Content
Causes and solutions
Содержание MagCore HF16Plus
Страница 1: ...Ver 2020 09 10 04 ...
Страница 8: ...7 IQ Documents IQ OQ PQ Documents ...
Страница 9: ...8 RBCBioscience OQ Documents ...
Страница 10: ...9 PQ Documents ...
Страница 34: ...33 Disassemblethebackservice lid ...
Страница 35: ...34 RBCBioscience Disassemblethetopservicelid ...
Страница 36: ...35 Axisdirections V axis direction X axis direction Y axis direction C axis direction M axis direction ...
Страница 46: ...45 ReplaceDoor Sensor 1 Open the front door 2 Pull out the sensor 3 Disconnect the connector ...
Страница 49: ...48 RBCBioscience ReplaceCoolingFans 3 Loosen two screws and remove protecting net ...
Страница 62: ...61 ReplaceTouchPanel 1 Open the front door 2 Loosen two screws 3 Remove fixed block 4 Remove the HMI ...
Страница 64: ...63 Electrical control and engineering software ...
Страница 68: ...67 New Type PLC configuration ...
Страница 77: ...76 RBCBioscience PositionTeaching Tools 200µlSPTip toolA toolB toolC PipetteTip ...
Страница 89: ...88 RBCBioscience ...
Страница 95: ...94 RBCBioscience ...
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......