56
RBCBioscience
Replace the lift motor of the base
Open the front door and take out the waste box, cartridge rack and T-rack.
1. Unplug the power connector to the lift motor.
2. Screw off the six screws on the liquid container above the heating module (2mm
Hex wrench).
3. Screw off the two screws on the metal sheet of the motor (3mm Hex wrench).
4. Screw off the two screws on the timing gear (2.5mm Hex wrench).
5. Screw off the four fixation screws on the motor (2.5mm Hex wrench).
6. Replace a new lift motor.
7. Install the new lift motor on the machine.
8. Note that the belt should be a little tight after installation.
Содержание MagCore HF16Plus
Страница 1: ...Ver 2020 09 10 04 ...
Страница 8: ...7 IQ Documents IQ OQ PQ Documents ...
Страница 9: ...8 RBCBioscience OQ Documents ...
Страница 10: ...9 PQ Documents ...
Страница 34: ...33 Disassemblethebackservice lid ...
Страница 35: ...34 RBCBioscience Disassemblethetopservicelid ...
Страница 36: ...35 Axisdirections V axis direction X axis direction Y axis direction C axis direction M axis direction ...
Страница 46: ...45 ReplaceDoor Sensor 1 Open the front door 2 Pull out the sensor 3 Disconnect the connector ...
Страница 49: ...48 RBCBioscience ReplaceCoolingFans 3 Loosen two screws and remove protecting net ...
Страница 62: ...61 ReplaceTouchPanel 1 Open the front door 2 Loosen two screws 3 Remove fixed block 4 Remove the HMI ...
Страница 64: ...63 Electrical control and engineering software ...
Страница 68: ...67 New Type PLC configuration ...
Страница 77: ...76 RBCBioscience PositionTeaching Tools 200µlSPTip toolA toolB toolC PipetteTip ...
Страница 89: ...88 RBCBioscience ...
Страница 95: ...94 RBCBioscience ...
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......