19
━ If the test result of the
optic measurement is over
detection
Select to enable or disable the optic measurement
function. (MagCore Super Only)
( 201, 202, 211, 105 without the function)
Please confirm the parameters and press the Start
button to execute the program.
Automatic extraction process.
The extraction process is completed.
If you select the optic measurement funtion, you
can browse, print or output the test result to a USB
Flash Drive (please refer to the optic data chapter).
f
A.Please record and save the
current measured data.
Please select YES.
Please check the value of each sample. If
there is a value showing DNA >300 ng/
μl or RNA > 240 ng/μl, please record
the number of the sample. It means the
sample needs to be diluted. After recording
the number of the slot, please insert the
USB Flash Drive and press the Save button
to save to current measured data. After
finishing the above steps, please press Next
Step.
10
7
8
9
1
The values are DNA >300 ng/μl /
RNA > 240 ng/μl
Please save the data in the USB Flash Drive
and transfer to computer, or the data in
the USB Flash Drive may be overwritten
later.
File name: OpticsSampleData.cvs
(MagCore Super Only)
Содержание MagCore HF16Plus
Страница 1: ...Ver 2020 09 10 04 ...
Страница 8: ...7 IQ Documents IQ OQ PQ Documents ...
Страница 9: ...8 RBCBioscience OQ Documents ...
Страница 10: ...9 PQ Documents ...
Страница 34: ...33 Disassemblethebackservice lid ...
Страница 35: ...34 RBCBioscience Disassemblethetopservicelid ...
Страница 36: ...35 Axisdirections V axis direction X axis direction Y axis direction C axis direction M axis direction ...
Страница 46: ...45 ReplaceDoor Sensor 1 Open the front door 2 Pull out the sensor 3 Disconnect the connector ...
Страница 49: ...48 RBCBioscience ReplaceCoolingFans 3 Loosen two screws and remove protecting net ...
Страница 62: ...61 ReplaceTouchPanel 1 Open the front door 2 Loosen two screws 3 Remove fixed block 4 Remove the HMI ...
Страница 64: ...63 Electrical control and engineering software ...
Страница 68: ...67 New Type PLC configuration ...
Страница 77: ...76 RBCBioscience PositionTeaching Tools 200µlSPTip toolA toolB toolC PipetteTip ...
Страница 89: ...88 RBCBioscience ...
Страница 95: ...94 RBCBioscience ...
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......