![RayTek RAYCMLTJM Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/raytek/raycmltjm/raycmltjm_operating-instructions-manual_3970115028.webp)
Accessories
CM
Rev. B 06/2011
28
7.6
Protective Window
The protective window comes with Silicon as window material.
Determination of transmissivity of an unknown protective window:
If transmissivity of the measuring screen is not indicated on the data sheet,
you can also determine the transmissivity yourself. Please proceed as follows:
1.
Measure the temperature of the target object with the sensing head,
without using the protective window. Note correct setting of
emissivity.
2.
Insert the protective window in the sensing head.
3.
Adjust the transmissivity in the software until the same temperature is
displayed, as it was determined without the protective window.