Technical Data
11
Rev. B 06/2011
CM
3
Technical Data
3.1
Parameters
Temperature range
-20 to 500°C (-4 to 932°F)
Spectral response
8 to 14 μm
Thermal Parameters
Accuracy
1
(Digital and Voltage)
± 1.5% of reading or ± 2°C
2
whichever is greater
Accuracy
3
(TC)
± 1.5% of reading ±2°C or ± 4°C
4
whichever is greater
Repeatability (Digital and Voltage)
± 0.5% of reading or ± 1°C
whichever is greater
Repeatability (TC)
± 0.5% of reading ± 1°C or ± 2°C
whichever is greater
Response time (95%)
150 ms
Temperature resolution
0.1°C (0.2°F)
Emissivity
0.100 to 1.100 (software controlled)
Transmissivity
0.100 to 1.000 (software controlled)
Electrical Parameters
Power
24 VDC ± 20% @ 20 mA
Analog Output
0 to 5 V or TCJ or TCK output
Digital Output
Two-way RS232 digital output
9600 baud, 8 data bits, 1 stop bit,
no parity, no flow control
1
for ambient temperature 23°C (73°F) ± 5 K , e = 0.95 and calibration geometry
2
± 3.5°C for T
meas
< 0°C (32°F)
3
for ambient temperature 23°C (73°F) ± 5 K , e = 0.95 and calibration geometry
4
± 5.5°C for T
meas
< 0°C (32°F)