Accessories
27
Rev. B 06/2011
CM
7.5
Right Angle Mirror
The Right Angle Mirror is used to turn the field of view by 90° against the
sensor axis. It is recommended when space limitations or excessive radiation
do not allow for direct alignment of the sensor to the target. The mirror must
be installed after the bracket and after the Air Purge Collar and screwed in
fully. In dusty or contaminated environments, air purging is required to keep
the mirror surface clean.
Figure 10: Dimension of Right Angle Mirror
When using the Right Angle Mirror, adjust the emissivity or
transmissivity settings downward by 5%. For example, for an
object with an emissivity of 0.65, you adjust the value down to
0.62. Or, you can keep the emissivity 0.65 and adjust the
transmissivity from 1.0 to 0.95. This correction accounts for
energy losses in the mirror.