Technical Data
MP150
13
2.3
Electrical Specifications
Communications
Ethernet
TCP/IP protocol 10/100 MBit/s, electrically isolated,
auto-negotiation
RS485
full-duplex, not addressable
9.6, 57.6, 115.2, 230.4 kBaud
Outputs
Analog
3 active current outputs, each adjustable 0/4 ... 20 mA,
maximum load: 500
Ω,
minimal load: 47
Ω
common ground connection for all current outputs, electrically
isolated to the GND ground
Alarm
potential-free relay contacts: 30 V / 1 A
normally open / normally closed
Inputs
Trigger
+ 5 to 24 VDC pulse, high/low active
Functional Input
max. + 5 VDC
function depends on the specific MP150 based system
Signal Processing
Stand-alone unit
Max, Min, Average, Peak/Valley Hold, Alarm setpoints
further signal processing function through software
Power Supply
24 VDC ± 25%, 1 A
2.4
General Specifications
Mechanical Scanning System
40,000 h (about 4.5 years) meantime between failures (MTBF)
for the mechanical scanning system at 48 Hz scan rate
Environment Rating
IP65 (NEMA 4) in accordance to IEC 60529
Ambient Temperatures
0 to 50°C (32 to 122°F) – without water cooling
max. 180°C (356°F) – with water cooling
down to –40°C (–40°F) – only for scanner with internal heating
element (optional) and still air (e.g. by using a protective
housing)
Storage Temperatures
-25 to 65°C (-13 to 149°F)
Water Cooling
max. pressure 15 bar (218 PSI)
Air Purge Collar
max. pressure 3 bar (43 psig)
Warm-up Time
30 min.
Relative Humidity
10 to 90%, non-condensing,
for operating and non-operating temperature range
Vibration
IEC 60068-2-6, 3 axes, 10 to 150 Hz, operating 2 g above 20 Hz
Shock
IEC 60068-2-27, 3 axes, operating: 5 g at 11 ms, 15 g at 6 ms
Dimensions