Technical Data
8
MP150
2
Technical Data
2.1
Measurement Specifications
Temperature Range
LT
20 to 350°C (68 to 662°F)
MT
100 to 800°C (212 to 1472°F)
HR
100 to 650°C (212 to 1202°F)
G5
100 to 950°C (212 to 1742°F)
P3
30 to 250°C (86 to 482°F) – for P30
100 to 350°C (212 to 662°F) – for P31
2M
400 to 950°C (752 to 1742°F)
1M
600 to 1200°C (1112 to 2192°F)
Spectral Response
LT
3 to 5 µm
MT
3.9 µm
HR
3.5 to 4 µm
G5
5 µm
P3
3.43 µm
2M
1.6 µm
1M
1 µm
Detectors
LT, MT, HR, G5, P3
MCT
2M
InGaAs
1M
Si
System Accuracy
LT
± 2°C (± 4°F)
MT
± 0.5% of reading or ± 3°C (6°F), whichever is greater
HR
± 0.5% of reading or ± 3°C (6°F), whichever is greater
G5
± 0.5% of reading or ± 3°C (6°F), whichever is greater
P3
± 3°C (± 6°F)
1M, 2M
± 0.5% of reading or ± 3°C (6°F), whichever is greater
Repeatability
LT
± 1°C (± 2°F)
MT
± 1°C (± 2°F)
HR
± 1°C (± 2°F)
G5
± 1°C (± 2°F)
P3
± 1°C (± 2°F)
1M, 2M
± 2°C (± 4°F)
1
at 0 - 50°C (32 - 122°F) ambient temperature
2
at 0 - 50°C (32 - 122°F) ambient temperature