Programming
Guide
Description
Char Format
P B S Legal values
Factory
default
LCD
Presel. setpoint / relay
function
SV
nnn.n
(1)
Target temperature
T nnn.n
* *
in current scale
(°C / °F)
Temperature
unit
U X
* * * C
/
F
C
U
Poll / Burst mode
V X
*
* P = poll
B = burst
Poll mode
Burst string contents
X$
*
Multidrop address
XA nnn
*
* 000 – 032
000 = single unit
000
Bottom temperature of
range
XB nnn.n
*
Restore factory defaults
XF
*
Transmission
XG n.nnn
* * * 0.100
-
1.000 1.000 T
High temperature of
range
XH nnn.n
*
Sensor initialization
XI n
* * * 1 after Reset,
0 if XI = 0
Box temperature
XJ nnn.n
* *
in current
scale(°C / °F)
FTC 3 trigger / hold
XN X
*
* T = trigger
H = hold
T
Analog output mode
XO n
*
* 0 = 0 – 20 mA
4 = 4 – 20 mA
5 = TCJ
6 = TCK
9 = mV
9
Firmware revision
XR
*
e.g. 1.01
Setpoint / relay function
(10)
XS nnn.n
* * (1)
250°C
Trigger
XT n
* *
0 = inactive
1 = active
0
Unit identification
XU
*
e.g. !XUMILT4
Serial number
XV
*
e.g. 98123
Advance hold hysterese
(4)
XY nnn.n
* *
Head calibration (9)
XZ
*
*
(1)
LT:
‐
40
to
600°C
(
‐
40
to
1112°F)
(2)
n
=
number,
X
=
uppercase
letter
78
MID
Содержание 54301
Страница 56: ...Accessories 8 2 Adjustable Mounting Bracket Figure 31 Adjustable Mounting Bracket XXXMIACAB 48 MID...
Страница 57: ...Accessories 8 3 Fixed Mounting Bracket Figure 32 Fixed Mounting Bracket XXXMIACFB MID 49...
Страница 63: ...Accessories Figure 39 Air Cooling System T Adapter MID 55...
Страница 64: ...Accessories Hose inner 9 mm 0 35 in outer 12 mm 0 47 in Figure 40 Dimensions of Air Cooling System 56 MID...
Страница 66: ...Accessories 8 7 Box Lid Figure 43 Box Lid with View Port for Post Installations XXXMIACV 58 MID...