Programming
Guide
11.2
General
Command
Structure
Requesting
a
parameter
(Poll
Mode)
?ECR
“?“
is
the
command
for
“Request“
“E“
is
the
parameter
requested
“CR“
(carriage
return,
0Dh)
is
closing
the
request.
Remark:
It
is
possible
to
close
with
“CR“
“LF“,
0Dh,
0Ah,
but
not
necessary.
Setting
a
parameter
(Poll
Mode)
The
parameter
will
be
stored
into
the
device
EEPROM.
E=0.975CR
“E“
is
the
parameter
to
be
set
“=“
is
the
command
for
“set
a
parameter“
“0.975“
is
the
value
for
the
parameter
“CR“
(carriage
return,
0Dh)
is
closing
the
request
Remark:
It
is
possible
to
close
with
“CR“
“LF“,
0Dh,
0Ah,
but
not
necessary.
Setting
a
parameter
without
writing
into
the
EEPROM
(Poll
Mode)
This
function
is
for
test
purposes
only.
E#0.975CR
“E“
is
the
parameter
to
be
set
“#“
is
the
command
for
“set
parameter
without
writing
into
the
EEPROM“
“0.975“
is
the
value
for
the
parameter
“CR“
(carriage
return,
0Dh)
is
closing
the
request.
Remark:
It
is
possible
to
close
with
“CR“
“LF“,
0Dh,
0Ah,
but
not
necessary.
Device
response
format:
!E0.975CRLF
“!“
is
the
parameter
for
“Answer“
“E“
is
the
parameter
“0.975“
is
the
value
for
the
parameter
“CR“
„LF“
(0Dh
0Ah)
is
closing
the
answer.
MID
67
Содержание 54301
Страница 56: ...Accessories 8 2 Adjustable Mounting Bracket Figure 31 Adjustable Mounting Bracket XXXMIACAB 48 MID...
Страница 57: ...Accessories 8 3 Fixed Mounting Bracket Figure 32 Fixed Mounting Bracket XXXMIACFB MID 49...
Страница 63: ...Accessories Figure 39 Air Cooling System T Adapter MID 55...
Страница 64: ...Accessories Hose inner 9 mm 0 35 in outer 12 mm 0 47 in Figure 40 Dimensions of Air Cooling System 56 MID...
Страница 66: ...Accessories 8 7 Box Lid Figure 43 Box Lid with View Port for Post Installations XXXMIACV 58 MID...