Installation
5.3.2
Head
Ambient
Temp.
/
Alarm
Output
This
output
can
be
configured
either
as
output
for
the
head
ambient
temperature
(default
configuration)
or
as
an
alarm
output.
Power +
Power –
Head Ambient Temp.
Ground
Figure
10:
Wiring
the
Output
for
Head
Ambient
Temperature
The
output
range
for
the
head
ambient
temperature
is
0
to
500°C
(32
to
932°F)
with
10
mV
/°C.
In
case
of
an
alarm
the
output
switches
between
0
V
and
5
V.
The
alarm
output
is
controlled
by
the
target
temperature
or
the
sensing
head
temperature.
Power -
Power +
Figure
11:
Wiring
of
the
Alarm
Output