1.4.
Testing Methodology
1.4.1.
Pre-Charge
In order to stabilize the testing battery, a pre-charge step is introduced at the start of all steps in each
methodology automatically. The system will apply a small charging current (200mA) for 5 minutes.
1.4.2.
Waken Charge
Small current rate is used to charge the battery pack. It will not check for the output current since some batteries
will not allow any current flow during the initial stage. The Waken Recover will be self terminated in programmed
time.
Step description:
Step
Type
Mode
Set Value
Current Output
Terminate Conditions
1
Charge
CV
4V/cell
200mA
Programmed time
1.4.3.
Storage Charge
Single charge step to charge the battery pack to normal operational status.
Step description:
Step
Type
Mode
Set Value
Current Output
Terminate Conditions
1
Charge
CC
0.5 C
0.5C
Volt > 4.1V/cell
1.4.4.
High Charge
Single charge step to charge the battery pack to above 93% charged state.
Step description:
Step
Type
Mode
Set Value
Current Output
Terminate Conditions
1
Charge
CV
0.5 C
0.5C
Current < 0.02C
1.4.5.
Discharge
Single discharge step to discharge the battery pack to near empty state, user can terminate the test at any time to
achieve required battery level.
Step description:
Step
Type
Mode
Set Value
Current Output
Terminate Conditions
1
Discharge
CC
0.5 C
0.5C
Volt < 3.0V/cell