4.3.
Testing information display N/A
When the RSOC always show N/A when the channel is under testing:
1.
Refer to Section 4.2 to correct the SMBus Read
2.
If the N/A shows occasionally, ignore the RSOC reading, the test
can still be carried out
When all the read cell voltages are showing N/A:
1.
Refer to Section 4.2 to correct the SMBus Read
2.
Older version of the battery packs employed older version of
SMBus Gas Gauge ICs. These older Gas Gauge ICs do not
support cell voltages in their register data. When tried Section
4.2 and all other SMBus data can be read correct, most likely the
battery does not support cell voltage. Ignore the reading and
continue with the test scheme
4.4.
Test abort abnormally
When test is forced to abort without any user interaction, check the
first LCD display screen and the second line will show the reason of
abort.
1.
Suspects loosen connection between battery and the channel.
Check the contacts between the connector and the battery,
some batteries have smaller contact points that some PIN in the
connector might not make the full contact.
2.
Go into the Last Step Data screen by turning the Dial Button,
check the battery voltage, if the voltage is 0V, the battery might
have been protected. Try Charge or Discharge the battery
according to the status.
3.
Go into Cell Voltage screen by turning the Dial Button, check
individual cell voltage to see if there is any imbalance among the
cells. Over-voltage and Under-voltage protections will be
triggered when one of the cell voltage exceed the protection
limit.
4.
Over Voltage protected during charging test: the battery pack
might goes into over voltage protection when charged if it is
already high charged (RSOC > 93%), Try to discharge it for 30
minutes and restart the charging test.
Storage N/A%
> > > > > > > >
Read RSOC Error
N/AV N/AV
N/AV
Read Cell Voltage Error
Storage Abort
Current 0A
11.35V 0.00A
1850mAH 01:20:30
3.78V 3.77V
3.80V