www.ravas.com
31
Parameter
Function
Settings
Default
US
01
Start-up unit (and print units)
1=kg / 2=lb
2
02
Smallest graduation step for
multi-range
0.1/0.2/0.5…………10/20/50
0.5
03
Largest graduation step for
multi-range
0.1/0.2/0.5 ……….. 10/20/50
2
04
Number of graduations for
every range
0000-9900 divisions
1000
05
Weighing capacity system (full
scale)
0000-99999 units
5000
06
Motion tolerance for stable
1
0.5 grad./sec
2 grad./sec
8 grad./sec
0-32
off
1 grad./sec
4 grad./sec
16 grad./sec
32 grad./sec
07
Filter size
8
0-12
0=off
1=light filtering, 12=heavy filtering
08
Auto zero range
0=off
0.5 division
0.5
1=division
3 divisions
09
Zero range positive (+)
0-100% (approved 2%) of span
10
10
Zero range negative (-)
0-100% (approved 2%) of span
10
11
Test Function
ADC Counts
bASIC
BASIC
10x Resolution
12
Power On zero
Yes / No
No
13
Approved
None, NTEP, OIML, NTEPC
None
14
Start-up number to add in
sampling mode
1-2-5-10-20-50-95
10
15
Units switch mode active
Yes / No
No
16
Setpoint function
0-14
0 (not used), 1 (gross overload), 2 (net overload), 3
(fill manual tare), 4 (fill auto tare), 5 (gross overload
not authorize to change gross settin), 6 (net overload
not authorize to change net setting), 7(gross overload
delayed), 8 (net overload delayed), 9 (gross overload
delay not authorize to change gross setting), 10 (net
overload delay not authorize to change net
setting),11 (gross oerrors), 12(gross
oerr, no changes allowed), 13(gross
oerr, delayed), 14(gross oerr, no
changes allowed, delayed)
0