The
orientation
can
be
started
from
the
Calibration
window
on the
Start screen
.
●
place
the
calibration
plate
at the working
distance to
the
scanner (the first calibration
position),
●
press
the
Capture
button,
●
next,
following
the
steps,
make
two more
shots.
●
after
this
the
accuracy
of orientation will be
shown. It should be approximately equal to the
accuracy value
of
the
last
calibration.
If
the
accuracy
during
orienting
starts
to
be very different
from
the
original
one,
you
need
to
re-calibrate
the
equipment.
Example
:
The
initial
accuracy
is
0.084,
accuracy
after
orientation
is
0.25.
The
value
exceeded
0.2,
re-calibration
is
required.
Automatic
calibration
The
automatic
calibration
can
be
started
from
the
Calibration
window
on the
Start
screen
, but
it
is
only
accessible
when
turntable
is
connected and recognized by the
software.
Automatic
calibration
serves
the
same
purpose
as
full
calibration,
but
it is easier to
perform, because it
uses
turntable
to
shift
the
calibration plate between
positions.