![Rangevision NEO Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/rangevision/neo/neo_setup-and-calibration_3643667013.webp)
After
all
the
necessary
snapshots
are
taken,
wait
for
the
result
of
calibration procedure.
From now on the
resulting accuracy
of
calibration
will be
specified on the home screen
in
pixels
.
A
good
result is accuracy
no worse (not higher) than 0.2 pix
. If you get a higher
result, repeat the calibration, carefully checking the plates position and brightness. If
is
is
necessary, repeat calibration without
blue
light
and
filters.
Some
advice
on
calibration:
●
do
not
change
the
distance
from
the
center
of
the
plate
to
the
scanner
when
turning
the
plate.
The
exception
are
only
snapshots
in
positions
10
and
11,
●
do
not
rotate
the
plate
at
a
very
high
angle.
Ensure
that
all
marks
on
the
plate
are
visible
from
both
cameras,
when
you
turn
the
table,
●
be
careful
with
calibration
plates!
It
is
not
permitted
to
contaminate
or
perform
any
mechanical
damage
to
the
surface
with
the
marks.
After
the
use
store
the
plates
in
designated
case,
●
before
the
calibration
procedure
ensure
that
cameras
and
cables
are
secure
in
the
designated
position.