VNA GUI reference
R&S
®
ZNL/ZNLE
441
User Manual 1178.5966.02 ─ 19
Ripple Check
Switches the ripple limit check of the active trace on or off. When the limit check is
switched on, a movable info field shows the pass/fail information and the measured rip-
ple in each ripple limit range.
Limit violations are marked with colored squares.
An
and a
indicating pass or fail can be generated in addi-
tion.
Note:
●
Ripple check and display of limit lines are independent of each other:
–
The ripple limits can be displayed, no matter if the ripple check is enabled.
–
If "Ripple Check" is enabled, the ripple limits are checked, no matter if they are
displayed.
–
The ripple check can even be enabled, if no limit lines are defined. In this case,
the info field displays "No ripple defined!" and the limit check always passes.
●
For each trace, ripple limits can only be set for a single cartesian trace format. If
another trace format is selected, the ripple limit lines are hidden and the ripple
check is suspended.
Remote command:
CALCulate<Chn>:RIPPle:STATe
CALCulate<Chn>:RIPPle:FAIL?
CALCulate:RIPPle:FAIL:ALL?
CALCulate<Chn>:RIPPle:SEGMent<Seg>:RESult?
CALCulate<Chn>:RIPPle:STATe:AREA
Ripple Fail Beep
Activates or deactivates the fail beep. The fail beep is a low-tone acoustic signal that is
generated each time the analyzer detects an exceeded ripple limit. No fail beep can be
generated if the ripple limit check is switched off.
Note:
The R&S
ZNL/ZNLE does not have a built-in audio device and loudspeaker. To
hear sounds, connect a USB audio device to the instrument or operate it via remote
desktop.
Remote command:
CALCulate<Chn>:RIPPle:SOUNd[:STATe]