Troubleshooting
R&S
®
TSMA6
98
User Manual 4900.8057.02 ─ 06
Table 8-2: Software license key information
Label
Description
Option Type
Band or technology option name
Option Material No.
Order number of the option
Option key
Software license key number
Privilege
Usage type (customer, services, demo)
Time Stamp
Time the software license key was installed
License Count
Number of times the (band upgrade) option is installed
Activation Type
Activation can be permanent or temporary
Valid From
Start of validity for temporary license
Valid To
End of validity for temporary license
Time to Expiration
Time left until license expires
Option Index
for internal use only
Format ID
for internal use only
8.4
R&S
TSMA6 Automatically Switches Off After Power
On
1. Check if there is a battery pack unit connected.
● If no, contact R&S support.
2. Check the state of charge of the batteries.
● If the charging state is > 20% (charging indicator displays one loading bar),
continue with
● If the charging state is < 20%, an auto power off of the device is performed due
to insufficient remaining battery charge. Continue with
3. Perform one of the following actions.
a) Replace the batteries with fully charged ones and try again.
b) Charge the batteries in an external charger and retry later.
c) Connect DC supply to the R&S
TSMA6 battery pack unit and charge the batter-
ies inside the bay. Keep the R&S
TSMA6 switched off.
4. Connect DC supply to the R&S
TSMA6 battery pack unit and check in the web GUI
the "Shutdown Settings".
Navigate to "Configuration" > "System" > "Power" and check the state of "Auto
Power Off".
● If enabled, check the shutdown delay.
R&S