Instrument Tour
R&S
®
TSMA6
15
User Manual 4900.8057.02 ─ 06
The LAN interface can be used for Remote Control / Remote Desktop Connections.
Alternatively, it can also be used for distributed applications of R&S NESTOR.
The LEDs on the LAN port indicate the status of the connection to the host PC. LED 1
(yellow, link status) is on the left side of the port, LED 2 (green, activity status) is on the
right.
HDMI Port
The HDMI port can be used for connecting an external monitor. (max. resolution: 2560
x 1600 pixel)
RF IN Connector
The RF IN connector is the RF input of the R&S
TSMA6. The multi-band RF paddle
antenna (700 MHz to 2.6 GHz), which is included in the shipment of R&S
TSMA6 or
any custom RF source is connected to this SMA connector. The maximum input power
is +20
dBm/10
V
DC.
Risk of instrument damage
Do not overload the maximum allowed input of +20
dBm.
Non-compliance destroys the input mixer.
Battery Pack Connector
The battery pack connector is used to connect a battery pack unit (R&S
TSMA6-BP,
R&S No. 4900.9001.02) allowing mobile operation.
3.3
Status LEDs
The four LEDs on the rear panel display the following states.
●
[Scanner State / Scanner Pwr]
Indicates the state of the scanner component
●
[Mode]
Indicates the device status
●
[Meas]
Indicates the state of the SW application
Status LEDs