Instrument Control
R&S
®
SGS100A
46
Getting Started 1416.0570.02 ─ 11
Using the PCIe interface for remote control of the instrument requires
extended knowledge and is described in section "Advanced Remote Con-
trol via PCIe" in the user manual.
5.2.1.4
GPIB Interface (IEC/IEEE Bus Interface)
The R&S
SGS is not equipped with an IEC/IEEE bus interface.
To be able to control the instrument via the GPIB bus:
1. Connect a GPIB-to-LAN or a GPIB-to-USB adapter to the instrument.
2. Use a GPIB bus cable to connect the instrument and the controller.
3. Provide the GPIB bus card, the card drivers and the program libraries for the
programming language in the controller.
4. In the "SGMA-GUI > Setup > Instruments > instrument name > Remote Con-
trol", set the "GPIB Address".
5. If the controller is equipped with several GPIB bus cards, define the used
"Board Number".
GPIB address
The controller must address the instrument with the GPIB bus channel. GPIB pro-
vides channel addresses from 0 to 30.
The GPIB resource string is
GPIB::<address>[::INSTR]
, where:
●
GPIB
denotes the used interface
●
<address>
indicates the used channel
●
[::INSTR]
indicates the instrument resource class (optional)
Note:
If the VISA implementation supports the GPIB interface, you can optionally
define the VISA instrument control resource (INSTR). It is used to define the
basic operations and attributes for a device, such as reading, writing, or trigger-
ing.
Any connected IEC bus cable must be terminated by an instrument or con-
troller.
Network and Remote Control Operation