First Steps with the Instrument
R&S
®
SGS100A
36
Getting Started 1416.0570.02 ─ 11
In the R&S SGS100A Web-GUI main panel, the green indicator "System Sta-
tus > OK " confirms that there is a connection between the instrument and the
remote PC and that the instrument is recognized by the software.
If you want to get additional information about the instrument, click on the
"Device Name". For additional information on other settings, hold the mouse
cursor over the specific setting.
Error messages are also displayed in the R&S SGS100A Web-GUI. If you
want to hide an error message, click on it.
4. In the menu bar on the left side press "Preset" to restore the predefined instru-
ment's settings.
5. Select "Frequency" and adjust the setting as required.
6. Select "RF On" to enable the output of the CW signal.
A signal with the set frequency is output at the [RF OUT] connector at the rear
panel of the R&S
SGS.
Configuring a CW Signal with the R&S SGS100A Web-GUI