Preparing for use
R&S
®
QAR50
14
User Manual 1179.4992.02 ─ 02
4.3
Installing the R&S
QAR50
The R&S
QAR50 is a large and heavy device. Unpacking, moving and installing the
R&S
QAR50 requires special arrangements. Follow the instructions provided here
when you install the R&S
QAR50.
See also:
●
●
"Lifting, carrying and installing the product"
Parts of delivery required for installation
●
1 x R&S
QAR5050
●
2 x supporting feet, incl. screws (M8X45) and washers
●
16 x mounting hole caps (M8X16 screws)
●
1 x power cable
●
2 x Allen wrench (1x size 5 and 1 x size 6)
Other required tools for installation
●
Open-ended wrench size 19
●
Spirit level
Setting up the R&S
QAR50
You can set up the R&S
QAR50 in two different ways: lying on its long side ("lying posi-
tion") or standing upright on one of its short sides ("upright position"). By default, the
R&S
QAR50 is designed to stand on a pair of feet that are part of the delivery.
For the upright position, make sure that the side with the connector board is at the bot-
tom.
The R&S
QAR50 must be placed on a level surface. This is important not only for
safety reasons, but also to get valid measurement results. If the body of the
R&S
QAR50 is not at a right angle to the floor, measurement results can be invalid.
Installing the R&S