5.2.3
Measurements
For bumper measurements, the R&S
QAR provides a reflection measurement. Bumper
measurements require the optional reference reflector. The measurement usually con-
sist of two steps, normalization and the actual measurement.
Measuring a DUT
1. Set up the normalization.
2. Start the normalization with "Normalize".
For details on how to normalize the measurement, see
3. Place the DUT between the R&S
QAR and the reference reflector.
4. Start the reflection measurement with "Measure" (or the F5 key).
The measurement itself lasts less than a second. The results are displayed after a
few seconds.
For details on how to run the measurement, see
.......................................................................................49
5.2.3.1
Normalization
Normalization is a measurement done in the free space, to obtain a reference value for
the bumper measurements. You do not need any object to measure or additional ele-
ment.
Normalization enables you also to check the positioning and the status of the reference
reflector. If the position is wrong, or the reflector is damaged, you get a warning.
Perform the following steps:
1. Make sure that nothing is obstructing the view between R&S
QAR and the refer-
ence reflector.
2. Choose the configuration needed from the "bumper type" drop-down menu.
3. Click the "Normalize" button to start the measurement.
The R&S
QAR measures the deviations and saves the correction data. All subse-
quent measurements are corrected accordingly.
4. Optional: scan the free space by clicking the "Measure" button to verify the normali-
zation.
Bumper Measurements