Remote Control Commands
R&S
®
NRX
328
User Manual 1178.5566.02 ─ 05
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
-290.0 to +110.0
*RST:
+0.0
Default unit: dBm
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:DIRection
<direction>
Selects the direction of forward power relative to the defined ports
1 and 2.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<direction>
AUTO | FORWard | REVerse
AUTO
The port with the greater measured power is interpreted as for-
ward power.
*RST:
AUTO
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:PEP:HOLD:TIME
<value>
Sets the hold time of the peak hold circuit of the power sensor.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
1e-3 to 1e-1
*RST:
0.01
Default unit: s
Manual operation:
See
Configuring Sensors