Remote Control Commands
R&S
®
NRX
221
User Manual 1178.5566.02 ─ 05
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<ref>
Range:
-290.0 to +110.0
*RST:
+30.0
Default unit: dBm
Manual operation:
See
14.4.6
Trace
Further information:
●
14.4.6.1
Trace Measurement Settings
In a trace measurement, you can use commands that combine several setting com-
mands. They are described in
Chapter 14.4.6.2, "Combining Trace Commands"
The same principle is used for the calculation functions, see
CALCulate<Measurement>:TRACe:X:POINts
................................................................... 221
[SENSe<Sensor>:]BWIDth:VIDeo:LIST?
.......................................................................... 222
[SENSe<Sensor>:]BANDwidth:VIDeo:LIST?
.....................................................................222
[SENSe<Sensor>:]TRACe:MEASurement:AUTO[:STATe]
...................................................222
[SENSe<Sensor>:]TRACe:MEASurement[:STATe]
............................................................ 222
[SENSe<Sensor>:]TRACe:REALtime
...............................................................................222
CALCulate<Measurement>:TRACe:X:POINts
<points>
Sets the number of required values per trace sequence.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<points>
Range:
1 to 8192
*RST:
660
Measurement Settings and Results