Measurement Types and Result Displays
R&S
®
NRX
64
User Manual 1178.5566.02 ─ 05
The measurement result is a single scalar value, either an absolute value or related to
a reference value.
8.1.2
Continuous Average Settings
Access: "Measurement Settings"
> "Measurement Type"
> "Continuous Average"
............................................................................................. 64
.................................................................................. 64
Unit
See
Resolution
See
Display
See
Chapter 7.1, "Display Settings"
Rel
Groups the settings for relative measurements.
Reference Value ← Rel
Available if
is set to "On" or "Set".
Sets the reference value.
Remote command:
[SENSe<Sensor>:]POWer:REFerence
CALCulate<Measurement>:RELative<DirectionalChannel>:POWer[:
MAGNitude]
CALCulate<Measurement>:RELative<DirectionalChannel>:RATio[:
MAGNitude]
Relative Measurements ← Rel
Allows you to relate measured power to a reference value. Whether the power is mea-
sured by one power sensor or whether it is a combined value measured by two power
sensors is set by
"Channel Calculation Function"
Off
Displays the absolute power or power ratio.
On
Displays the relative power or power ratio. As reference value, the
value specified under
is used.
Set
Assigns the current measurement result as reference value and dis-
plays the relative power.
Continuous Average