Measurement Types and Result Displays
R&S
®
NRX
102
User Manual 1178.5566.02 ─ 05
Standing Wave Ratio (SWR) ← Reflection
1 - Reflection coefficient
Standing wave ratio =
1 + Reflection coefficient
See also
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"POWer:SWRatio"
Return Loss ← Reflection
Reverse power
Return loss =
Forward power
10 x log
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"POWer:RLOSs"
Reflection Coefficient ← Reflection
Reflection coefficient =
Forward power
Reverse power
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"POWer:RCOefficient"
Reflection Ratio ← Reflection
Forward power
Reflection ratio =
Reverse power
100
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"POWer:RFRatio"
Display
See
Chapter 7.1, "Display Settings"
Autoscale
Adapts the scaling of the graphical display.
Trigger
Opens the "Measurement Trigger Configuration" dialog:
●
●