Remote Control Interfaces and Protocols
R&S
®
NRQ6
54
Getting Started 1178.3705.02 ─ 07
The IP address or hostname is used by the programs to identify and control the
sensor. While the hostname is determined by settings in the sensor, the IP
address is assigned by a DHCP server when the sensor requests one. Alterna-
tively the IP address is determined with a procedure called Zeroconf.
You can also assign a
LAN device name
which defines the protocol characteris-
tics of the connection. See the description of the VISA resource string below for
the corresponding interface protocols. The string of the
LAN device name
is
emphasized in italics.
VXI-11
TCPIP::<IP address or hostname>
[::inst0]
[::INSTR]
●
inst0
is the LAN device name, indicating that the VXI-11 protocol is used
(optional)
inst0
currently selects the VXI-11 protocol by default and can be omitted.
Chapter 7.2.2, "VXI-11 Protocol"
HiSLIP
TCPIP::<IP address or hostname>
::hislip0
[::INSTR]
●
hislip0
is the HiSLIP device name, designates that the interface protocol
HiSLIP is used (mandatory)
hislip0
is composed of [::HiSLIP device name[,HiSLIP port]] and must be
assigned.
Chapter 7.2.3, "HiSLIP Protocol"
Socket communication
TCPIP::<IP address or hostname>
::port::SOCKET
●
port
determines the used port number
●
SOCKET
indicates the raw network socket resource class
Socket communication requires the specification of the port (commonly referred
to as port number) and of "SOCKET" to complete the VISA resource string with
the associated protocol used.
The default port for socket communication is port 5025.
Chapter 7.2.4, "Socket Communication"
Ethernet Interface
Содержание NRQ6
Страница 1: ...R S NRQ6 Frequency Selective Power Sensor Getting Started Getting Started Version 07 1178370502 ÜU52 ...
Страница 6: ...Contents R S NRQ6 6 Getting Started 1178 3705 02 07 ...
Страница 8: ...Safety Information R S NRQ6 8 Getting Started 1178 3705 02 07 ...
Страница 14: ...Key Features R S NRQ6 14 Getting Started 1178 3705 02 07 ...
Страница 30: ...Preparing for Use R S NRQ6 30 Getting Started 1178 3705 02 07 Connecting to a Controlling Host ...
Страница 38: ...R S NRQ6 Tour R S NRQ6 38 Getting Started 1178 3705 02 07 Fan Openings ...
Страница 50: ...Operating Concepts R S NRQ6 50 Getting Started 1178 3705 02 07 Remote Control ...
Страница 58: ...Contacting Customer Support R S NRQ6 58 Getting Started 1178 3705 02 07 ...