R&S
NRQ6 Tour
R&S
®
NRQ6
32
Getting Started 1178.3705.02 ─ 07
5.1
RF Connector
See (1) in
.
The male N connector is used to connect the R&S
NRQ6 to the device under test
(DUT) or a signal generator, see
Chapter 4.4, "Connecting to a DUT"
For maximum measurement accuracy, tighten the RF connector using a torque
wrench with the recommended nominal torque. For details, see
Risk of overloading the sensor
Using the R&S
NRQ6 at a level above its upper measuring limit can dam-
age the sensor head. To avoid this risk, make sure not to exceed the test
limit.
5.2
Status Information
See (2) in
.
The status LED shows the state of the R&S
NRQ6 by color and blinking fre-
quency.
Table 5-1: Possible states
Color
Illumination
State
White
Steady
Idle
The sensor performs no measurement and is
ready for use.
White
Blinking
Firmware update is in progress.
Yellow
Steady
Waiting for trigger state.
Green
Steady
Measurement is running.
Turquoise blue Steady
Zeroing is in progress.
Status Information
Содержание NRQ6
Страница 1: ...R S NRQ6 Frequency Selective Power Sensor Getting Started Getting Started Version 07 1178370502 ÜU52 ...
Страница 6: ...Contents R S NRQ6 6 Getting Started 1178 3705 02 07 ...
Страница 8: ...Safety Information R S NRQ6 8 Getting Started 1178 3705 02 07 ...
Страница 14: ...Key Features R S NRQ6 14 Getting Started 1178 3705 02 07 ...
Страница 30: ...Preparing for Use R S NRQ6 30 Getting Started 1178 3705 02 07 Connecting to a Controlling Host ...
Страница 38: ...R S NRQ6 Tour R S NRQ6 38 Getting Started 1178 3705 02 07 Fan Openings ...
Страница 50: ...Operating Concepts R S NRQ6 50 Getting Started 1178 3705 02 07 Remote Control ...
Страница 58: ...Contacting Customer Support R S NRQ6 58 Getting Started 1178 3705 02 07 ...