Preparing for Use
R&S
®
NRQ6
17
Getting Started 1178.3705.02 ─ 07
4.4
Connecting to a DUT
The R&S
NRQ6 has fan openings at both sides of the casing, as shown in
. When connecting the R&S
NRQ6 to a DUT and setting up the measure-
ment assembly, be careful to allow sufficient airflow.
Risk of damage due to insufficient airflow
If the R&S
NRQ6 is run with insufficient airflow for a longer period, the
R&S
NRQ6 overheats. Overheating can disturb the operation, even cause
damage and lead to wrong measurement results.
Make sure of the following:
●
All fan openings are unobstructed.
●
Airflow perforations are unimpeded.
●
Minimum distance between the fan openings and any object is 10
cm.
To connect to the DUT
For connecting the R&S
NRQ6 to a DUT, use the RF connector of the
R&S
NRQ6. For details, see
1. Ensure that the RF connector of your DUT is compatible with the RF connec-
tor of the R&S
NRQ6.
2. Insert the RF connector straight into the RF output of your DUT. Take care not
to tilt it.
SMART SENSOR TECHNOLOGY
NRQ6
Frequency Selective Power Sensor
LAN PoE+
Host
Interface
Sample Clock I/O (Default: 120 MHz)
LO I/0
180
3.
NOTICE!
Risk of damaging the center pin of the RF connector. Only rotate the
hex nut of the RF connector. Never rotate the R&S
NRQ6 itself.
Tighten the RF connector manually.
4. Tighten the RF connector using a torque wrench with the nominal torque of
1.36
Nm (12''
lbs) to ensure maximum measurement accuracy.
Connecting to a DUT
Содержание NRQ6
Страница 1: ...R S NRQ6 Frequency Selective Power Sensor Getting Started Getting Started Version 07 1178370502 ÜU52 ...
Страница 6: ...Contents R S NRQ6 6 Getting Started 1178 3705 02 07 ...
Страница 8: ...Safety Information R S NRQ6 8 Getting Started 1178 3705 02 07 ...
Страница 14: ...Key Features R S NRQ6 14 Getting Started 1178 3705 02 07 ...
Страница 30: ...Preparing for Use R S NRQ6 30 Getting Started 1178 3705 02 07 Connecting to a Controlling Host ...
Страница 38: ...R S NRQ6 Tour R S NRQ6 38 Getting Started 1178 3705 02 07 Fan Openings ...
Страница 50: ...Operating Concepts R S NRQ6 50 Getting Started 1178 3705 02 07 Remote Control ...
Страница 58: ...Contacting Customer Support R S NRQ6 58 Getting Started 1178 3705 02 07 ...