Operating Concepts
R&S
®
NRPxxS(N)
31
User Manual 1177.5079.02 ─ 10
Setup
Ethernet Switch
(PoE)
1
4
6
5
2
3
HOST
INTERFACE
IN: 3 V or 5 V logic
OUT: min. 2 V into 50
Ω
max. 5.3 V
TRIG2
I/0
PoE
SMART SENSOR TECHNOLOGY
NRP
7
Figure 5-1: Setup with the web user interface
1
= Signal source
2
= LAN power sensor
3
= RJ-45 Ethernet connector
4, 6 = RJ-45 Ethernet cable
5
= Ethernet switch supporting PoE power delivery
7
= Computer with a supported web browser installed
Incorrectly connecting/disconnecting the R&S NRPxxS(N) power sensors can damage
the power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensor as described in
Starting a measurement
1. Connect the cables as shown in
For a detailed description, refer to
Chapter 3.5.2, "Setting Up a LAN Connection"
2. Open a supported web browser.
3. Enter the instrument name or the IP address of the sensor you want to connect to.
Example: http://nrp33sn-123456
For details on how to find out the IP address or hostname, refer to
on page 21 and
Chapter 3.5.2.3, "Using Hostnames"
Browser-Based User Interface