Remote Control Commands
R&S
®
NRPxxS(N)
132
User Manual 1177.5079.02 ─ 10
Suffix:
<Channel>
.
1...4
Measurement channel if more than one channel is available.
Usage:
Query only
SYSTem:ERRor:ALL?
Queries all unread entries in the error/event queue and removes them from the queue.
The response is a comma-separated list in first out order, each entry consisting of the
error number and a short description of the error.
Positive error numbers are instrument-dependent. Negative error numbers are
reserved by the SCPI standard.
Example:
SYST:ERR:ALL?
Query
0,"No error"
Response
Usage:
Query only
SYSTem:ERRor:CODE:ALL?
Queries all unread entries in the error/event queue and removes them from the queue.
Only the error numbers are returned.
Example:
SYST:ERR:CODE:ALL?
Query
0
Response: No errors have occurred since the error queue was
last read out.
Usage:
Query only
SYSTem:ERRor:CODE[:NEXT]?
Queries the oldest entry in the error queue and then deletes it. Only the error number
is returned.
Example:
SYST:ERR:CODE?
Query
0
Response: No errors have occurred since the error queue was
last read out.
Usage:
Query only
SYSTem:ERRor:COUNt?
Queries the number of entries in the error queue.