Browser-Based User Interface
R&S
®
NRPxxS(N)
50
User Manual 1177.5079.02 ─ 10
A value of 0.0 corresponds to an ideal matched source. A value of 1.0 corresponds to
total reflection.
Remote command:
[SENSe<Sensor>:]SGAMma:MAGNitude
Phase ← Γ Correction
Sets the phase angle of the complex reflection coefficient of the source,
Γ
source
.
Remote command:
Zero Calibration
Performs zeroing using the signal at the sensor input. See
Note:
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
Remote command:
CALibration<Channel>:ZERO:AUTO
Diagnostics
Triggers a selftest of the sensor.
Note:
Do not apply a signal to the sensor while the selftest is running. If the selftest is carried
out with a signal being present, error messages can erroneously be output for the test
steps
Offset Voltages
and/or
Noise Voltages
.
The results of the selftest are shown in a dialog that is displayed after the test comple-
tion.
Settings